| Message |
Meaning |
Remedy |
|
CAL
|
Calibration data error
|
Service
|
|
CHR
|
Charge controller data error
|
Charge
|
|
CNF
|
Configuration is wrong
|
Repeat configuration
|
|
FIL
|
File system error
|
Clear data file
|
|
NSR
|
Configuration memory error
|
Service
|
|
OPT
|
Configuration error
|
Service
|
|
TIM
|
Time administration error
|
Set time and date
|
|
TVL intern active
|
This warning hint (TCA=temperature sensor for flue gas) is displayed, if no external sensor is plugged in or if the instrument can´t recognize it. The instrument switches to an internal temperature sensor, so that the stack loss and the efficiency may be measured without an external temperature sensor. The internal temperature sensor has not the accuracy and the response time of the external sensor and doesn´t meet the demands of the European Standard.
|
|
|
TG missing/fault
|
This warning hint (TG = temperature sensor for flue gas) is displayed if the thermocouple of the probe cannot be detected, because it is missing (probe has no thermocouple) or because the thermocouple or the plug are faulty. All values which need this temperature, are displayed as "---"
|
|
|
O2-Sensor
|
Probe has been in flue
Sensor defect
|
Calibrate again with fresh air
Service
|
|
CO-Sensor
|
Probe has been in flue
Sensor defect
|
Calibrate with fresh air
Service
|
|
NO-Sensor
|
NO-Sensor has been without current
Others like CO-sensors
|
Check function
See CO-sensor
|
|
NO2-Sensor
|
Probe has been in flue
Sensor defect
|
Calibrate with fresh air
Service
|
|
SO2-Sensor
|
Probe has been in flue
Sensor defect
|
Calibrate with fresh air
Service
|
|
Pressure Sensor
|
Sensor with pressure during calibration
Sensor defect
|
Calibrate without pressure
Service
|
|
System Temperature
|
Operating temperature range exceeded
|
Vary temperature
Service
|
|
Junction Temperature
|
Operating temperature range exceeded
|
Vary temperature
Service
|
|
Battery Temperature
|
Operating temperature range exceeded
|
Vary temperature
Service
|
|
Battery Voltage
|
Battery voltage not between 4,4 V and 6,5 V
|
Replace battery
Service
|
|
Battery current
|
Battery current for charging or operating too high
|
Service
|
|
Battery Error
|
Battery manager data error
|
Charge battery
|
|
Set Clock
|
Clock lost data
|
Charge battery
Set clock
|
|
Gas Flow
|
Gas flow out of operating range
|
Check gas duct
Service
|
|
Options
|
Data record options error
|
Service
|
|
Settings
|
Wrong setting
|
Check and change settings
|
|
Calibration Data
|
Error in calibration data record
|
Service
|
|
Data Memory
|
Error in memory data record
|
Service
|
|
Next Service
|
Next service data error
|
Service
|
|
Fuel table
|
Fuel table error
|
Reload fuel table with pc-program
|
|
Display Table
|
Fuel table error
|
Reload display table with pc-program
|
|
Printer Table
|
Printer table error
|
Reload printer table with pc-program
|
|
System Configuration
|
System configuration error
|
Service
|